COSMOSIL HPLC Columns

Silica Based Preparative Columns

COSMOSIL series is available in 10.0 mm I.D. and 20.0 mm I.D. for semi-preparative applications and 28.0 mm I.D. and 50.0 mm I.D. for preparative scales. For column sizes and packing materials not listed below, contact either your local distributor or the manufacturer directly.

Material characteristics

Packing material 15C18-MS-II 15C18-AR-II 15C18-PAQ
Silica gel high purity porous spherical silica
Average particle size 15 m
Average pore size approx. 120
Specific surface area approx. 300 m2/g
Stationary phase octadecyl group
Bonding type monomeric polymeric
Main interaction hydrophobic interaction
End capping treatment near-perfect treatment
Carbon content approx. 16% approx. 17% approx. 11%
pH range 2-10* 1.5-7.5* 2-7.5
Feature This phase is recommended for most of applications but particularly effective for basic organic compounds. This phase is recommended for the separations requiring acidic mobile phase conditions. It also shows superior molecular shape selectivity to monomeric type C18columns. This phase is designed to offer superior retention of polar compounds and excellent reproducibility in highly aqueous mobile phases, even in 100% aqueous.

* Optimum pH range of columns based on silica gel is between 2 and 7.5.

Comparison of particle size

When change particle size of packing material from 5 m to 15 m, the number of theoretical plate (N) is reduced by one-third, and the pressure is reduced by one-ninth. As shown in the figures below, when a small amount of sample is injected, there is a big difference in the number of theoretical plates between 5 m and 15 m. However, when a large amount of sample is injected, there is not much difference between the two. Therefore the low pressure packing material (particle size 15 m) is recommended for preparative columns (28 mm I.D. or more).

The number of theoretical plate of N=4. Naphthalene peak.

Comparison of particle size

Condition
Column size 4.6 mm I.D. x 250 mm Detection UV 254 nm
Mobile phase methanol : water = 70 : 30 Pressure 5C18-AR-II/ 10.9 MPa
15C18-AR-II/1.8 MPa
Flow rate 1.0 ml/min Sample
  1. Uracil
  2. Methyl Benzoate
  3. Toluene
  4. Naphthalene
Temperature 30°C

Preparative separation using 50.0 mm I.D. column

Carbazol is extracted from anthracene oil (coal tar) and required high purity because it is often used for analytical applications. Following is the preparative separation of carbazol using a 50.0 mm I.D. COSMOSIL 15C18-AR-Ⅱ .

Preparative separation using 50.0 mm I.D. column

Condition
Column COSMOSIL 5C18-AR-II
4.6 mm I.D. x 150 mm
Flow rate 0.5 ml/min
Temperature room temperature
Detection UV 254 nm
Condition
Column COSMOSIL 15C18-AR-II
50 mm I.D. x 250 mm
Flow rate 60 ml/min
Temperature room temperature
Detection UV 254 nm

Ordering information

Preparative column (Particle size : 15 m)
COSMOSIL 15C18-MS-II
Product name Size Product No.
COSMOSIL 15C18-MS-II Guard Column
28 mm l.D. x 50 mm 05885-51
50 mm l.D. x 50 mm 34527-41
COSMOSIL 15C18-MS-II Packed Column
28 mm l.D. x 250 mm 34525-61
50 mm l.D. x 250 mm 05886-41
50 mm l.D. x 500 mm 34531-71
COSMOSIL 15C18-AR-II
Product name Size Product No.
COSMOSIL 15C18-AR-II Guard Column
28 mm l.D. x 50 mm 38030-11
50 mm l.D. x 50 mm 38057-81
COSMOSIL 5CN-MS Packed Column
28 mm l.D. x 250 mm 37978-51
50 mm l.D. x 250 mm 38058-71
50 mm l.D. x 500 mm 05884-61
COSMOSIL 15C18-PAQ
Product name Size Product No.
COSMOSIL 15C18-PAQ Guard Column
28 mm l.D. x 50 mm 05887-31
50 mm l.D. x 50 mm 05889-11
COSMOSIL 15C18-PAQ Packed Column
28 mm l.D. x 250 mm 05888-21
50 mm l.D. x 250 mm 05890-71
50 mm l.D. x 500 mm 05891-61

▲Page top