COSMOSIL NPE

COSMOSIL NPE column is a reversed phase column with nitrophenylethyl groups bonded silica packing material. This column provides unique retention characteristics, slightly different from the COSMOSIL PYE column, utilizing both dipole-dipole and π-π interactions.

Selectivity for Dipole

Selectivity for dipole is evaluated based on the separation of 1,5-dinitronaphthalene and 1,8-dinitronaphthalene. Dinitronaphthalenes (peak 1 and 2) were strongly retained on PYE and NPE because of π-π interaction compared with dimethylnaphthalenes. However, there is a slight difference between these two columns. While 1,5-dinitronaphthalene (peak 2) was preferentially retained on PYE, 1,8-dinitronaphthalene (peak 1) was retained longer on NPE. The results with NPE indicate the presence of strong dipole-dipole interaction. The two nitro group dipoles in 1,8-dinitronaphthalene are aligned for a much greater dipolar coupling with the bonded nitrophenyl group in NPE than 1,5-dinitronaphthalene.

data

Condition
Column Size 4.6 mm I.D. x 150 mm structures
Mobile Phase 5NPE Methanol : Water = 70 : 30
5PYE Methanol : Water = 90 : 10
5C18-MS-II Methanol : Water = 80 : 20
Flow Rate 1.0 ml/min
Temperature 30°C
Detection UV 254 nm
Sample
  1. 1,8-Dinitronaphthalene 0.21 µg
  2. 1,5-Dinitronaphthalene 0.11 µg
  3. Naphthalene 0.25 µg
  4. 1-Methylnaphthalene 0.35 µg
  5. 1,5-Dimethylnaphthalene 0.42 µg
Attention
  1. Methanol is recommended as a mobile phase for COSMOSIL NPE column. Acetonitrile is not recommended because it has many π electrons and interferes π-π interactions between a sample and the stationary phase.
  2. The stationary phase of COSMOSIL NPE, nitrophenyl group, has a large UV absorption. When the stationary phase detaches from silica gel and elutes, even a slight quantity can be detected and causes baseline noise. In such a case, wash the column with tetrahydrofuran. Detachment of a small amount of the stationary phase does not deteriorate a column?s separation ability.
  3. COSMOSIL NPE column is not suitable for gradient analysis.

Ordering Information

Analytical Column (Particle Size : 5 µm)
COSMOSIL 5NPE
Product Name Size Product No.
COSMOSIL 5NPE Guard Column
4.6 mm I.D. x 10 mm 37904-01
10 mm I.D. x 20 mm 38045-31
20 mm I.D. x 20 mm 05868-81
20 mm I.D. x 50 mm 05869-71
COSMOSIL 5NPE Packed Column
1.0 mm I.D. x 150 mm 05897-01
2.0 mm I.D. x 150 mm 34328-51
2.0 mm I.D. x 250 mm 34379-91
4.6 mm I.D. x 150 mm 37902-21
4.6 mm I.D. x 250 mm 37990-71
10 mm I.D. x 250 mm 05469-11
20 mm I.D. x 250 mm 38046-21