・ Excellent for
Basic compounds
・ Improved separation for acidic or metal coordination compounds
・ Simple mobile phase conditions
Conventionally end-capped C18 columns still have many residual silanol groups on the silica surface of stationary phase that can form ionic bonds with basic compounds. The resulting peak tailing makes accurate quantification of analytes difficult, especially in trace analyses. COSMOSIL 3C18-EB with a better end-capping treatment offers improved peak shape and separation particularly for basic compounds.
| COSMOSIL 3C18-EB | Competitor A C18, 3 µm |
Competitor B C18, 3 µm |
Competitor C C18, 3 µm |
|---|---|---|---|
![]() |
![]() |
![]() |
![]() |
| Condition | ||||
|---|---|---|---|---|
| Column Size | 4.6 mm I.D. x 75 mm | |||
| Mobile Phase | Acetonitrile/ 20mmol/l Phosphate Buffer(pH7.0) = 60/40 |
|||
| Flow Rate | 1.0 ml/min | |||
| Temperature | 40°C | |||
| Detection | UV 254 nm | |||
| Sample |
|
|||
| Inj. Vol | 2.0 µl | |||
| COSMOSIL 3C18-EB | Competitor A C18, 3 µm |
Competitor B C18, 3 µm |
Competitor C C18, 3 µm |
|---|---|---|---|
![]() |
![]() |
![]() |
![]() |
| Condition | ||||
|---|---|---|---|---|
| Column Size | 4.6 mm I.D. x 75 mm | |||
| Mobile Phase | Acetonitrile/ 20mmol/l Phosphate Buffer(pH7.0) = 60/40 |
|||
| Flow Rate | 1.0 ml/min | |||
| Temperature | 40°C | |||
| Detection | UV 234 nm | |||
| Sample |
|
|||
| Inj. Vol | 1.5 µl | |||
Acidic or metal coordination compounds can also form bonds with residual silanol groups and metal impurities causing peak tailing. COSMOSIL 3C18-EB with a better end-capping treatment and high purity silica gel offers improved peak shape and separation for acidic or metal coordination compounds.
| COSMOSIL 3C18-EB |
|---|
![]() |
| Condition | ||||
|---|---|---|---|---|
| Column Size | 4.6 mm I.D. x 75 mm | |||
| Mobile Phase | Acetonitrile/ 20mmol/l Phosphate Buffer(pH2.5) = 30/70 |
|||
| Flow Rate | 1.0 ml/min | |||
| Temperature | 40°C | |||
| Detection | UV254 nm | |||
| Sample |
| |||
| COSMOSIL 3C18-EB |
|---|
![]() |
| Condition | ||||
|---|---|---|---|---|
| Column Size | 4.6 mm I.D. x 75 mm | |||
| Mobile Phase | Acetonitrile/ 20mmol/l Phosphoric Acid = 10/90 |
|||
| Flow Rate | 1.0 ml/min | |||
| Temperature | 40°C | |||
| Detection | UV 254 nm | |||
| Sample |
|
|||
| Inj. Vol | 2.0 µl | |||
If you use insufficiently end-capped C18 columns for basic compound analyses, you may need to spend much time adjusting
mobile phase conditions. You may need
In addition, the complexity of the mobile phase often is detrimental to reproducibility. COSMOSIL 3C18-EB with a new end-capping treatment allows good peak shape and separation
using simple mobile phase conditions.
| Complicated Condition | Simple Condition |
|---|---|
| Methanol/ Acetonitrile/ 20mmol/l KH2PO4, 0.2%Triethylamine(pH3.0 with H3PO4) = 35/15/50(*) | Acetonitrile/ 20mmol/l Phosphate Buffer(pH2.5) = 30/70 |
![]() |
|
| (*): USP 38(1) Amlodipine and Benazepril Hydrochloride Capsules, Performance Tests | |
| Condition | |||
|---|---|---|---|
| Column Size | 4.6 mm I.D. x 100 mm | ||
| Flow Rate | 1.0 ml/min | ||
| Temperature | 40°C | ||
| Detection | UV 237 nm | ||
| Sample |
|
||
| Inj. Vol | 50 µl | ||
3C18-EB has excellent synthesis reproducibility and column-to-column reproducibility, so it is applicable in routine analysis such as quality control of drugs.
| COSMOSIL 3C18-EB |
|---|
![]() |
| Condition | |||
|---|---|---|---|
| Column Size | 4.6 mm I.D. x 150 mm | ||
| Mobile Phase | Acetonitrile / H2O = 60/40 | ||
| Flow Rate | 1.0 ml/min | ||
| Temperature | 40°C | ||
| Detection | UV 254 nm | ||
| Sample |
|
||
| Inj. Vol | 1.5 µl | ||
| COSMOSIL 3C18-EB |
|---|
![]() |
| Condition | |||
|---|---|---|---|
| Column Size | 4.6 mm I.D. x 150 mm | ||
| Mobile Phase | Acetonitrile/ 20mmol/l Phosphate Buffer(pH7.0) = 60/40 |
||
| Flow Rate | 1.0 ml/min | ||
| Temperature | 40°C | ||
| Detection | UV 254 nm | ||
| Sample |
| ||
| Inj. Vol | 2.0 µl | ||
COSMOSIL 3C18-EB produces sharp peaks using ammonium acetate buffer in LC-MS.
| Acetonitrile/ 20mmol/l Phosphate buffer(pH7.0) = 60/40 | Acetonitrile/ 10mmol/l CH3COONH4 = 60/40 |
|---|---|
![]() |
![]() |
| Condition | ||||
|---|---|---|---|---|
| Column Size | 4.6 mm I.D. x 75 mm | |||
| Flow Rate | 1.0 ml/min | |||
| Temperature | 40°C | |||
| Detection | UV 254 nm | |||
| Sample |
|
|||
| Inj. Vol | 1.0 µl | |||
| COSMOSIL 3C18-EB |
|---|
![]() |
| Condition | |||
|---|---|---|---|
| Column Size | 4.6 mm I.D. x 100 mm | ||
| Mobile Phase | A; Acetonitrile/ 36.7mmol/l Phosphate Buffer(pH7.6) = 10/90 B; Acetonitrile/ 36.7mmol/l Phosphate Buffer(pH7.6) = 80/20 B conc. 0→ 20%(0→ 10min), 20→ 100%(10→ 30min) |
||
| Flow Rate | 1.0 ml/min | ||
| Temperature | 40°C | ||
| Detection | UV302 nm | ||
| Sample |
| ||
| Inj. Vol: | 20 µl | ||
| USP 37(3) Esomeprazole Magnesium Delayed-Release Capsules, Impurities | |||
| Product Name | Size | Product No. |
|---|---|---|
| COSMOSIL 3C18-EB Guard Column | 4.6 mm I.D. x 10 mm | 09839-41 |
| COSMOSIL 3C18-EB Packed Column | 2.0 mm I.D. x 50 mm | 09794-21 |
| 2.0 mm I.D. x 75 mm | 09795-11 | |
| 2.0 mm I.D. x 100 mm | 09796-01 | |
| 2.0 mm I.D. x 150 mm | 09797-91 | |
| 2.0 mm I.D. x 250 mm | 09798-81 | |
| 3.0 mm I.D. x 50 mm | 09799-71 | |
| 3.0 mm I.D. x 75 mm | 09800-21 | |
| 3.0 mm I.D. x 100 mm | 09811-81 | |
| 3.0 mm I.D. x 150 mm | 09814-51 | |
| 3.0 mm I.D. x 250 mm | 09827-91 | |
| 4.6 mm I.D. x 50 mm | 09840-01 | |
| 4.6 mm I.D. x 75 mm | 09841-91 | |
| 4.6 mm I.D. x 100 mm | 09842-81 | |
| 4.6 mm I.D. x 150 mm | 09843-71 | |
| 4.6 mm I.D. x 250 mm | 09844-61 |