COSMO(+)Capillary is cationic polymer coated capillary column, for electroosmotic flow (EOF) is from the chthode (-) to the direction of the anode (+) because of the cationic polymer coating inside of the capillary column.

![]() |
Peak Identification: 1: Glycolate 2: Lactate 3: Fumarate 4: 2-Oxoisopentanoate 5: Succinate 6: Malate 7: PEP 8: DHAP 9: Glycerol3P 10: 3PG 11: MES(is) 12: Gluconate 13: Ribulose5P 14: G1P 15: G6P+F6P 16: 2,3DPG 17: Sedoheptulose7P 18, F1,6DP |
| Experimental Conditions: | |||
|---|---|---|---|
| Sample: Capillary: Electrolyte: Applied Potential: Injection: Temperature: Sheath liquid: |
E.coli W3110, 109 Cells COSMO(+) 50 µm I.D. x 110 cm 50 mM Ammonium Acetate, pH8.5 -30 kV 30 sec at 50 mbar 20°C 10 µl/min 5 mM Ammonium Acetate in 50% (v/v) Methanol - Water |
||
| Experimental Conditions: | |||
|---|---|---|---|
| Sample Concentration Capillary Electrolyte Applied Potential Injection Temperature Sheath liquid |
50 µmol/l each COSMO(+) 50 µm I.D. x 110 cm 50 mM Ammonium Acetate, pH8.5 -30 kV 30 sec at 50 mbar 20°C 10 µl/min 5 mM Ammonium Acetatein 50% (v/v) Methanol - Water |
||
![]() |
(a) Lactic Acid (m/z=89) (b) Succinic Acid (m/z=117) (c) Malic Acid (m/z=133) (d) Uric Acid (m/z=167) (e) Citrate (m/z=191) |
| Experimental Conditions: | ||||
|---|---|---|---|---|
| Capillary Electrolyte Applied Potential Injection Temperature Sheath liquid |
COSMO(+) 50 µm I.D. x 110 cm 50 mM ammonium acetate, pH8.5 -30 kV -10 kV × 30 sec 25°C 10 mM Ammonium Carbonate: 2-Propanol =1:1 |
|||
| Product name | PKG size | Product No. | |
|---|---|---|---|
| COSMO(+)Capillary (50 µm I.D. x 120 cm) | |||
| 2 PKG | 07584-44 | ||