Products

Poor peak shape

T1. Poor peak shape

Poor peak shapes

Symptom Cause Solution
Particular sample is tailing Undesirable ion exchange interaction between basic compounds and packing material. Use a column with fewer residual silanols (3C18- EB or 5C18-MS-II), or add 0.1-1% acid to mobile phase.
Undesirable coordinate interaction between metal coordination compound and packing material. Add 5 mmol/l of di-sodium dihydrogen ethylenediaminetetraacetate dihydrate (EDTA・2Na) to mobile phase.
Undesirable hydrogen bonding interaction between sample and packing material. Change the organic solvent (e.g., acetonitrile to methanol).
All samples are tailing Voids in packing material or column deterioration. Replace the column.
(If the tailing does not improve after replacing the column) Sample is dispersing inside the column. Reduce dead volume (refer to Q25 for more information on dead volume).
Fronting Injection of large volume of sample solvent that is significantly different in elution properties or pH compared to mobile phase. Dissolve sample in mobile phase. If the sample does not dissolve, dissolve in a soluble solvent first, then dilute with mobile phase.
Reduce injection volume to 1/2-1/10.
Caution: Spikes or peak broadening may also occur.
Broad peaks (1) Sample has high molecular weight (MW: 2,000 or greater). Proteins with high molecular weight cannot access pores of packing material. Use wide pore (pore size: 300Å) column for reversed phase chromatography, Protein-R. Please see web or catalog for more information.
Sample volume is too large. Reduce injection volume to 1/2-1/10.
Caution: Tailing peaks may also occur.
If a particular sample has a broad peak, the compound may be adsorbed to the packing material. Use Protein-R for a high recovery rate. Please see web or catalog for more information.
If all samples have broad peaks, the column may have deteriorated. Replace the column.
Concentration of ammonium sulfate in sample solution is too low on hydrophobic chromatography (HIC). Adjust concentration of ammonium sulfate to 1 mol/l or greater.
Broad peaks (2) Sample has low molecular weight (MW: 2,000 or less). Sample volume is too large. Reduce sample amount from 1/2 to 1/10.
Caution: Tailing peaks may occur instead of broad peaks.
If a particular sample has a broad peak, the compound may be adsorbed to the packing material. Replace with a column that has a different packing material.
(3C18-EB and 5C18-MS-II are recommended for basic samples. )
If all samples have broad peaks, the column may have deteriorated. Replace the column.
Peak spikes or peak splitting occur for certain samples. More than 2 samples are contained in the peak, and slightly separated. Find the conditions that separate the two samples.
Mobile phase and sample solvent are significantly different in their separation properties. Dissolve sample in mobile phase. If sample does not dissolve, dissolve sample in soluble solvent, then dilute with the mobile phase.
Reduce injection volume to 1/2 to 1/10.
Mixed dissociated and non-dissociated ionic sample. Adjust pH of mobile phase to pKa ±2 or more of the ionic sample.
Peak spikes or peak splitting occur for all samples Mobile phase and sample solvent are significantly different in their separation properties. Dissolve sample in mobile phase. If sample does not dissolve, dissolve sample in soluble solvent, then dilute with the mobile phase.
Reduce injection volume to 1/2 to 1/10.
The column may have deteriorated. Replace the column.

T2. Ghost peaks

Separation Mode Cause Solution
<Reversed Phase Chromatography>
Gradient elution
Peaks from water impurities Use new HPLC-grade distilled water.
Use a pre-column. Please refer to Technical Information, Baseline Noise in Gradient Elution (PDF 600KB)
<Reversed Phase Chromatography>
Protein samples
Sample may be adsorb to the column, and elute on the next analysis. Wash column, Please refer to Q3 for more information on washing methods.
COSMOSIL Protein-R, which has high recovery rate for protein separations, is recommended.
<All Separation Modes> Sample solvent and mobile phase are significantly different. Sample solvent has peaks. Dissolve sample in mobile phase. If sample does not dissolve, dissolve sample in a soluble solvent, then dilute with the mobile phase.
<All Separation Modes>
Peaks in blank analysis (Peak area decreases with each injection.)
Injector is contaminated. Wash by injecting with a syringe 20 ml of solvent (e.g., methanol) that can dissolve the contaminants.
Micro-syringe is contaminated. Wash with solvent (e.g., methanol, chloroform or water) to dissolve the pollutants. Ultrasonic cleaning is effective.
Others Contamination or deterioration of samples Prepare the sample again.
Stabilizers in the mobile phase Use HPLC-grade solvent without stabilizers.